4.3.4 Taking pressure applied at device as value
Example: performing position adjustment.
P01-xxxxxxxx-19-xx-xx-xx-032
The new value for the damping is now 30.0 s.
Jump to the next parameter with "E".
You can get back to the editing mode with "+" or
"–".
On-site display Operation
P01-xxxxxxxx-19-xx-xx-xx-158
The bottom line on the on-site display displays the pressure
present, here 3.9 mbar.
P01-xxxxxxxx-19-xx-xx-xx-159
Use "+" or "–" to switch to the "Confirm" option. The active
selection is highlighted in black.
P01-xxxxxxxx-19-xx-xx-xx-037
Use "E" to assign the value (3.9 mbar) to the POS. ZERO
ADJUST parameter. The device confirms the calibration
and jumps back to the parameter, here POS. ZERO
ADJUST (see next graphic).
P01-xxxxxxxx-19-xx-xx-xx-160
Switch to the next parameter with "E".