Fault Analogue output Possible reason Remedy
Measured value jumps to
lower values despite constant
level
Multiple echoes If possible, increase first echo factor
page 35
Optimise alignment
page 36
If necessary, remount Micropilot in a
less central position
Call Service
Measured value drops to lower
value on approach to 100%
Product in window
suppression zone
Match measuring range to window
suppression zone, page 38
For levels over 100 % change
E and F if necessary, see basic
calibration, page 22, 23
E641 appears when the tank is
filled or emptied
Echo too weak to
be detected
Change application parameter in
V0H3, page 39
Optimise alignment, page 36 or .
check mounting position,
page 9 - 11
if possible, select better position
Foam (measurement impossible)
➃☎Call Service
Analogue output does not
function correctly
Incorrect wiring or
wire broken
Check wiring
No power Connect power supply
Electronics fault
Call Service
order spare part
No smart communication
Incorrect wiring
EMC fault
Check wiring and min./max. load
see Technical Data, page 45
Check screening,
see Connection, page 13
Call Service
20 mA
4mA t
expected
happened
yes
yes
20 mA
4mA t
expected
happened
20 mA
E641
4mA t
expected
happened
yes
yes
no
yes
no
yes
no
yes
no
Chapter 7 Trouble-Shooting Micropilot FMR 231
38 Endress+Hauser