Gammapilot FMG50
Endress+Hauser 33
Shock resistance
IEC 60068-2-27; test Ea; 30 g, 18 ms, 3 shocks/direction/axis
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61326 series and NAMUR Recommendation EMC (NE 21). For details, please refer to the
Declaration of Conformity
4)
.
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART).
Maximum measured error during EMC testing: <0.5 % of the span.
Process conditions
General
• The measuring principle generally does not depend on the process conditions
• Take radiating media into consideration
The FHG65 gamma modulator must be used for radiating media. This does not apply for
concentration measurement with radiating media.
Process temperature
In the event of high process temperatures, ensure sufficient insulation between the process vessel
and detector (see -> "Ambient temperature"). If necessary, use the heat shield that is optionally
available.
Process pressure
Consider the influence of the pressure on the gas phase when calculating the necessary activity and
during adjustment.
Mechanical construction
Dimensions, weights Gammapilot FMG50
143×118
(5.63×4.65)
A
Ø
85 (3.35)
B
248 (9.76)
C
A0037984
4) Available for download at www.de.endress.com.