Heartbeat Verification Availability
Available for the following versions of feature 540 "Application package":
EH: Heartbeat Verification + Monitoring
Device functionality checked on demand
Verification of the correct functioning of the measuring device within specifications.
The verification result provides information about the condition of the device: Passed or Failed.
The results are documented in a verification report.
The automatically generated report supports the obligation to demonstrate compliance with
internal and external regulations, laws and standards.
Verification is possible without interrupting the process.
Advantages
No onsite presence is required to use the function.
The DTM
5)
triggers verification in the device and interprets the results. No specific knowledge is
required on the part of the user.
The verification report can be used to prove quality measures to a third party.
Heartbeat Verification can replace other maintenance tasks (e.g. periodic check) or extend the
test intervals.
5) DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare or a DTM-based process control system.