Gammapilot FMG50 HART Electrical connection
Endress+Hauser 17
5.
+
-
321
A0038895
 2 Connection terminals and ground terminal in the connection compartment
1 Internal ground terminal (to ground the cable shield)
2 Negative terminal
3 Positive terminal
Connect the cable.
6. Tighten the cable glands or cable entries so that they are leak-tight.
7. Screw the cover securely back onto the connection compartment.
8. Tighten the cover lock.
6.2.2 Connection with fieldbus connector
In the case of device versions with a fieldbus connector, the housing does not have to be
opened to establish the connection.
Pin assignment for connector M12-A
2
1
34
A0011175
Pin 1: Signal +
Pin 2: Not used
Pin 3: Signal –
Pin 4: Ground