Commissioning Gammapilot FMG50 HART
20 Endress+Hauser
2. Open the device in FieldCare or DeviceCare.
The dashboard (homepage) of the device is displayed:
Click "System -> Device management"
3. Reset the device in the "Device reset" parameter
The following types of reset can be selected:
Restart device
A soft reset is performed here. The device software performs all the diagnostics that would
also be performed by a hard reset by switching the device on/off.
Reset to factory default
It is always advisable to reset the customer parameters if you want to use a device with an
unknown history, or if the operating mode is changed. When a reset is performed, all
customer parameters are reset to the factory default values
Optional: reset to customer settings
If the device was ordered with a customized configuration, a reset restores these customer
settings configured at the factory.
A reset can also be performed on site via the operating keys (see the "Commissioning via
onsite operation" section).
7.2 Commissioning using the wizard
A wizard is provided in FieldCare or DeviceCare
1)
that guides the user through the initial
commissioning process.
1. Connect the device with FieldCare or DeviceCare.
1) FieldCare and DeviceCare are available for download at www.software-products.endress.com. To download the
software, it is necessary to register in the Endress+Hauser software portal.