Commissioning Gammapilot FMG50 HART
24 Endress+Hauser
Level basic calibration
Calibration time per calibration: 5 min!
1. Reset
Press both keys > 12 s
2. Start background calibration
Press both keys > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
3. Start empty calibration
Press the "Zero / 1" key > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
Wait 5 min until the green LED stops flashing
4. Start full calibration
Press the "Span / 2" key > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
Wait 5 min until the green LED stops flashing
A reset deletes all calibrations!
Status and power LED
A green LED that signals the status and button activation feedback is provided on the
electronic insert.
Behavior of the LED
The LED flashes once briefly when the device is started
When a key is pressed, the LED flashes to confirm the key activation
When a reset is performed, the LED flashes as long as both keys are pressed and the reset is
not yet active (countdown). The LED stops flashing once the reset is active.
The LED flashes while calibration is being performed via onsite operation