Gammapilot FMG50 Electrical connection
Endress+Hauser 33
5.9 FMG50 with RIA15
The RIA15 remote indicator can be ordered together with the device.
Product structure, feature 620 "Accessory enclosed":
Option PE "Remote indicator RIA15, non-hazardous area, aluminum field housing"
Option PF "Remote indicator RIA15, hazardous, aluminum field housing"
Alternatively available as an accessory, for details see Technical Information
TI01043K and Operating Instructions BA01170K
L
CAUTION
Pay attention to the Safety Instructions (XAs) when using the Gammapilot FMG50 with
the remote indicator RIA15 in hazardous environments:
XA01028R
XA01464K
XA01056K
XA01368K
XA01097K
Terminal assignment RIA15
+
Positive connection, current measurement
-
Negative connection, current measurement (without backlighting)
LED
Negative connection, current measurement (with backlighting)
Functional grounding: terminal in housing
The RIA15 process indicator is loop-powered and does not require any external power
supply.
The voltage drop to be taken into account is:
≤1 V in the standard version with 4 to 20 mA communication
≤1.9 V with HART communication
and an additional 2.9 V if display light is used
5.9.1 Connection of the HART device and RIA15 without
backlighting
Y
I
R
s
DC
1
2
3
LED
-
+
A0019567
 6 Block diagram of HART device with RIA15 process indicator without light
1 Device with HART communication
2 Power supply
3 HART resistor

Table of Contents