Gammapilot FMG50 Electrical connection
Endress+Hauser 35
Connection of the HART communication resistor module, RIA15 with backlighting
1
3
2
Y
I
R
s
DC
LED
-
+
A0020840
 9 Block diagram of HART device, RIA15 with light, HART communication resistor module
1 HART communication resistor module
2 Device with HART communication
3 Power supply
5.10 Wiring
L
CAUTION
Note the following before connecting:
If the device is used in hazardous areas, make sure to comply with national standards
and the specifications in the Safety Instructions (XAs). The specified cable gland must
be used.
The supply voltage must match the specifications on the nameplate.
Switch off the supply voltage before connecting the device.
Connect the potential matching line to the external ground terminal of the transmitter
before connecting the device.
Connect the protective ground to the protective ground terminal.
The cables must be adequately insulated, with due consideration given to the supply
voltage and the overvoltage category.
The connecting cables must offer adequate temperature stability, with due
consideration given to the ambient temperature.
1. Release the cover lock
2. Unscrew the cover
3. Guide the cables into the cable glands or cable entries
4. Connect the cables
5. Tighten the cable glands or cable entries so that they are leak-tight
6. Screw the cover securely back onto the connection compartment
7. Tighten the cover lock
Housing thread
The thread of the electronics and connection compartment is coated with lubricant
varnish.
Avoid additional lubrication.