Gammapilot FMG50 Operation
Endress+Hauser 45
6.3 Locking/unlocking configuration
6.3.1 Software locking
Locking via password in FieldCare / DeviceCare / SmartBlue App
Access to the configuration of the FMG50 can be locked by assigning a password. The
"User role" is set to "Maintainer" in the as-delivered state. The device can be fully configured
in the "Maintainer" role. Afterwards, access to the configuration can be locked by assigning
a password. The "User Role" is now set to "Operator". The configuration can be accessed by
entering the password.
The password is defined under:
System -> User management -> Define password
You can switch from the "Maintainer" to "Operator" user role under:
System -> User management -> Logout
Deactivating the lock via FieldCare / DeviceCare / SmartBlue App
After entering the password, you can enable the configuration of the FMG50 as an
"Operator" with the password. The "User role" then changes to "Maintainer"
Navigate to:
System -> User management -> Change user role
6.3.2 Hardware locking
Hardware locking can only be unlocked on the electronic insert (flip the switch). It is not
possible to unlock the hardware by communication.
6.4 Resetting to the default configuration
L
CAUTION
A reset may negatively impact the measurement. As a rule, a basic setup must be
performed again after a reset. All calibration data are deleted after a reset. A complete
recalibration is needed to put the measurement back into operation.
1. Connect the device with FieldCare or DeviceCare.
2. Open the device in FieldCare or DeviceCare.
The dashboard (homepage) of the device is displayed:
Click "System -> Device management"
3. Reset the device in the "Device reset" parameter