Commissioning Gammapilot FMG50
48 Endress+Hauser
6. Once all the pages have been completed, click "Finish" to close the Wizard.
If you cancel the Wizard before all the necessary parameters have been entered, the
device may be in an undefined state. In such situations, it is advisable to reset the
device to the factory default settings.
The following operating modes can be set via the Wizard:
Level
Min. or max. point level
Density measurement
Concentration measurement
Concentration measurement of radiating media
Configuration of gammagraphy detection: see Section 8.6
Recalibration of a density measurement: see Section 8.7
7.2.2 Device identification
The user guidance starts with the general configuration of the tag name and some HART
parameter settings.
A0042162
A0042163
7.2.3 Measurement settings
After this, the general "measurement settings" of the Gammapilot FMG50 can then be
made: