Commissioning Gammapilot FMG50
52 Endress+Hauser
The correlation between the current output and the calibration values is always linear in
the point level operating mode. In this sense, this operating mode is the same as the Level
operating mode with the "linear" type of linearization.
1. Selection: start with full calibration or start with empty calibration
Start calibration -> the calibration can be stopped once the pulse rate has
stabilized.
A0042168
2. Empty calibration point level: the radiation is switched on and the beam path is
completely free
If these conditions are met, empty calibration can be started.
A0042169
The empty calibration can be performed by pressing the "Start empty calibration"
button. The measurement then starts automatically and continues, at the very
maximum, for as long as has been configured for the calibration time. However,
the process can also be stopped manually by pressing the "Stop calibration"
button.
The calibration stops automatically as soon as a million pulses have been
totalized.
Alternatively, the empty calibration can also be entered directly.
For the "Next" button in the Wizard to be enabled, the value must, however, be
changed from the start value, at least temporarily.