EasyManua.ls Logo

Endress+Hauser Gammapilot FMG50 - Page 64

Endress+Hauser Gammapilot FMG50
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Gammapilot FMG50
64 Endress+Hauser
A0042213
The settings for the current output are then made in the "Output settings" step
Interface
In the Gammapilot FMG50, interface measurement is performed by measuring the
different densities of two media, such as oil and water. Interface measurement in a
calibration is therefore very similar to a multi-point density measurement with two density
calibration values.
I
r
min
r
r
max
I
min
I
max
A0042211
I Pulse rate
ρ Density
I
min
Minimum pulse rate
ρ
min
Minimum density, oil
I
max
Maximum pulse rate
ρ
max
Maximum density, water
Once the calibration points have been entered, the Gammapilot FMG50 calculates the
interface layer in % on its own. Here, 0% corresponds to the minimum density and 100%
to the maximum density.
The settings for the current output are then made in the "Output settings" step
Beam path length
The length of the beam path in the medium to be measured is specified here.
A0042201
Examples:

Table of Contents

Other manuals for Endress+Hauser Gammapilot FMG50

Related product manuals