Commissioning Gammapilot FMG50
66 Endress+Hauser
The settings for the current output are then made in the "Output settings" step
Concentration
In concentration measurements, linearization defines the correlation between the
measured density and the concentration.
The concentration measurement is therefore a density measurement with subsequent
linearization. The calibration process is identical to density measurement.
Linearization is performed on completion of the density calibration.
Example: take the necessary value pairs from the diagram.
C (%)
r
1,2
1,4
1,8 2,21,6 2,0
10
20
30
40
50
0
3
(g/cm )
A0042218
 19 Example of a linearization curve for concentration measurements
Linearization
Linearization table conditions
The table can consist of up to 32 "density value : concentration (%)" pairs
The table must decrease monotonically
The first value in the table must correspond to the minimum density value
The last value in the table must correspond to the maximum density value
1. Perform density calibration