Gammapilot FMG50 Commissioning
Endress+Hauser 71
Linearization table conditions
• The table can consist of up to 32 "concentration - linearized value" pairs.
• The table must decrease monotonically
• The first value in the table must correspond to the minimum concentration
• The last value in the table must correspond to the maximum concentration
The table values can be sorted as monotonically increasing using the "Table mode -> Sort
table" function.
Edit table: the index of the linearization point is entered in this field (1-32 points)
Customer input value: enter the normalized pulse rate
Customer value: concentration in %.
Activate table: the "Enable" option must first be selected before the linearization table is
used. The linearization table is not used as long as "Disable" is selected.
The linearization table can also be entered manually in the linearization module. This is
started by selecting the "Linearization" button:
A0042194
The normalized pulse rate and the customer value can be entered directly in table form in
this module.
The linearization table must be activated by selecting "Activate table" -> "Enable"