Gammapilot FMG50 Diagnostics and troubleshooting
Endress+Hauser 85
8.4 Diagnostic event
8.4.1 Diagnostic event in the operating tool
If a diagnostic event is present in the device, the status signal appears in the top left status
area of the operating tool along with the corresponding symbol for the event level in
accordance with NAMUR NE 107:
Failure (F)
Function check (C)
Out of specification (S)
Maintenance required (M)
Error-free operation: display is lit green
Alarm or warning: display is lit red
Calling up remedial measures
Navigate to the Diagnostics menu
In the Actual diagnostics parameter the diagnostic event is shown with event text
8.4.2 List of diagnostic events in the operating tool
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
007 Sensor defective Replace sensor electronics F Alarm
008 Sensor defective 1. Restart device
2. Contact service
F Alarm
062 Sensor connection faulty Check sensor connection F Alarm
064 Pulse rate out of range 1. Check process conditions
2. Check environmental
conditions
3. Replace device
C Warning
082 Data storage inconsistent 1. Check Data Unit
2. Contact service
F Alarm
Diagnostic of electronic
242 Firmware incompatible 1. Check software
2. Flash or change main
electronic module
F Alarm
252 Module incompatible 1. Check if correct electronic
modul is plugged
2. Replace electronic module
F Alarm
270 Main electronics
defective
Replace main electronics F Alarm
272 Main electronics faulty 1. Restart device
2. Contact service
F Alarm
273 Main electronics
defective
1. Emergency operation via
display
2. Change main electronics
F Alarm
282 Data storage inconsistent 1. Restart device
2. Contact service
F Alarm
283 Memory content
inconsistent
1. Transfer data or reset device
2. Contact service
F Alarm