Commissioning Gammapilot M FMG60 PROFIBUS PA
36 Endress+Hauser
Further options:
g/cm
3
g/l
lb/gal; [1g/cm
3
= 8.345 lb/gal]
lb/ft
3
; [1g/cm
3
= 62.428 lb/ft
3
]
°Brix; [1°Brix =270 (1 - 1/x)]
°Baumé; [1°Baumé = 144.3 (1 - 1/x)]
°API; [1°API = 131.5 (1.076/x - 1)]
°Twaddell; [1°Twaddell = 200 (x-1)]
"x" refers to the density in g/cm
3
. The formula indicates how many degrees this density
corresponds to.
6.4.7 "Min. density" (*07)
Local display
Min. density *07
0.9500 g/cm
3
This function is needed for density and concentration measurements only. It is used to specify
the lower limit of the density measuring range. The output current for this density is 4 mA.
6.4.8 "Max. density" (*08)
Local display
Max. density *08
1.2500 g/cm
3
This function is needed for density and concentration measurements only. It is used to specify
the upper limit of the density measuring range. The output current for this density is 20 mA.
6.4.9 "Pipe diameter unit" (*09)
Local display
Pipe diam. unit *09
mm
inch
This function is needed for density and concentration measurements only. It is used to select
the unit for the pipe diameter.
1 in = 25.4 mm