Gammapilot M FMG60 PROFIBUS PA Commissioning
Endress+Hauser 37
6.4.10 "Pipe diameter" (*0A)
Local display
Pipe diam. *0A
200 mm
This function is needed for density and concentration measurements only. It is used to specify
the irradiated measuring path L. With standard installation, this value is identical to the pipe
inner diameter D
I
. However, in the case of other installation variants (to extend the irradiated
measuring path), this value may be greater (see diagram). The pipe walls are not to be
considered a part of the measuring path.
1
1
1
L
L
L
a
D
I
a
L
30°
2.0 x D
I
A0018109
 4 Always specify the complete irradiated measuring path L in the "Pipe diameter" (*0A) function.
Depending on the installation, this value may be larger than the actual pipe diameter.
1 Gammapilot M
6.4.11 "Output damping" (*0B)
Local display
Output damping *0B
60 s