3.3.2 Required flow rate
The required flow rate depends on
• the ambient temperature at the water cooling jacket
• the inlet temperature
• the measuring range of the Gammapilot M
Typical values are given in the following tables:
Ambient temperature T
A
=75°C
Ambient temperature T
A
=100°C
Ambient temperature T
A
=120°C
3.4 Post-installation check
After installing the device, carry out the following checks:
• Is the device damaged (visual inspection)?
• Does the device correspond to the measuring point specifications for ambient temperature,
measuring range etc.?
• If available: Are the measuring point number and labeling correct (visual inspection)?
• Is the measuring device sufficiently protected against direct sunlight?
• Are the cable glands tightened correctly?
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 30 l/h 41 l/h 55 l/h 70 l/h 84 l/h
25 °C 30 l/h 30 l/h 30 l/h 45 l/h 61 l/h 77 l/h 93 l/h
30 °C 30 l/h 30 l/h 33 l/h 50 l/h 68 l/h 86 l/h 104 l/h
35 °C 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
40 °C 30 l/h 30 l/h 47 l/h 72 l/h 98 l/h 124 l/h 149 l/h
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
25 °C 30 l/h 30 l/h 42 l/h 64 l/h 87 l/h 110 l/h 133 l/h
30 °C 30 l/h 30 l/h 47 l/h 73 l/h 98 l/h 124 l/h 150 l/h
35 °C 30 l/h 30 l/h 54 l/h 84 l/h 113 l/h 143 l/h 173 l/h
40 °C 33 l/h 33 l/h 66 l/h 101 l/h 137 l/h 173 l/h 210 l/h
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 45 l/h 70 l/h 94 l/h 119 l/h 144 l/h
25 °C 30 l/h 30 l/h 50 l/h 77 l/h 104 l/h 131 l/h 158 l/h
30 °C 30 l/h 30 l/h 55 l/h 85 l/h 115 l/h 146 l/h 176 l/h
35 °C 32 l/h 32 l/h 64 l/h 98 l/h 133 l/h 168 l/h 203 l/h
40 °C 38 l/h 38 l/h 75 l/h 116 l/h 157 l/h 199 l/h 240 l/h