3.3.2 Required flow rate
The required flow rate depends on
the ambient temperature at the water cooling jacket
the inlet temperature
the measuring range of the Gammapilot M
Typical values are given in the following tables:
Ambient temperature T
A
=7C
Ambient temperature T
A
=10C
Ambient temperature T
A
=12C
3.4 Post-installation check
After installing the device, carry out the following checks:
Is the device damaged (visual inspection)?
Does the device correspond to the measuring point specifications for ambient temperature,
measuring range etc.?
If available: Are the measuring point number and labeling correct (visual inspection)?
Is the measuring device sufficiently protected against direct sunlight?
Are the cable glands tightened correctly?
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 30 l/h 41 l/h 55 l/h 70 l/h 84 l/h
25 °C 30 l/h 30 l/h 30 l/h 45 l/h 61 l/h 77 l/h 93 l/h
30 °C 30 l/h 30 l/h 33 l/h 50 l/h 68 l/h 86 l/h 104 l/h
35 °C 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
40 °C 30 l/h 30 l/h 47 l/h 72 l/h 98 l/h 124 l/h 149 l/h
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
25 °C 30 l/h 30 l/h 42 l/h 64 l/h 87 l/h 110 l/h 133 l/h
30 °C 30 l/h 30 l/h 47 l/h 73 l/h 98 l/h 124 l/h 150 l/h
35 °C 30 l/h 30 l/h 54 l/h 84 l/h 113 l/h 143 l/h 173 l/h
40 °C 33 l/h 33 l/h 66 l/h 101 l/h 137 l/h 173 l/h 210 l/h
Inlet temperature
Measuring range
50 mm 200 mm 400 mm 800 mm 1200 mm 1600 mm 2000 mm
20 °C 30 l/h 30 l/h 45 l/h 70 l/h 94 l/h 119 l/h 144 l/h
25 °C 30 l/h 30 l/h 50 l/h 77 l/h 104 l/h 131 l/h 158 l/h
30 °C 30 l/h 30 l/h 55 l/h 85 l/h 115 l/h 146 l/h 176 l/h
35 °C 32 l/h 32 l/h 64 l/h 98 l/h 133 l/h 168 l/h 203 l/h
40 °C 38 l/h 38 l/h 75 l/h 116 l/h 157 l/h 199 l/h 240 l/h