Gammapilot M FMG60 with FOUNDATION Fieldbus Wiring
Endress+Hauser 23
4.3 Terminal assignment
Terminal compartment A
L00-FMG60xxx-04-00-00-xx-002
Terminal compartment B
L00-FMG60xxx-04-00-00-xx-001
L1
L+
N
L-
+
-
1
2
(3)
(4)
14
15
90 … 253 VAC
18 … 36 VDC
+
-
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
5
6
7
8
9
10
11
12
+- - +
0
13
Terminal(s) Meaning
0 Grounding of the cable screen
1)
1) rated cross section > 1 mm
2
1, 2 Power supply
2)
2) rated cross section max. 2.5 mm
2
Compartment B:
3, 4
Compartment A:
(3)
1
,(4)
1
Signal output, depending on communication version:
• 4-20mA with HART
• PROFIBUS PA
• Foundation Fieldbus
(Depending on the device version ordered, the signal output is in connection compartment A or B,
see below)
!
Note!
For the versions of the Gammapilot M with fieldbus plug connectors (M12 or 7/8"), the signal
output is wired in compartment B on delivery and connected to the fieldbus plug connector (see
below, section "Fieldbus plug connectors"). In this case, the housing needs not to be opened for
connecting the signal line.
5, 6 Pulse input (for cascading mode; is used for master and slave)
7, 8 Pulse output (for cascading mode; is used for slave and end slave)
9, 10, 11, 12 PT-100 input (4-wire)
13 Plug for display and operating module VU331 (normally in FHX40);
is wired on delivery and connected to the FHX40 plug
14 Protective earth
1
15 Protective earth or grounding of the cable screen
1