Operation Gammapilot M FMG60 with FOUNDATION Fieldbus
38 Endress+Hauser
5.5 Reset to the default configuration
It is advisable to reset the customer parameters if you want to use a device with an unknown history.
Effects of resetting:
• All customer parameters are reset to their default values.
• Linearization is switched to "linear", but the table values are kept. The table can be switched
back on in the "linearization" (*4) function group in the "linearization" (*40/*46) function.
In order to carry out the reset, enter "33333" in the "reset" (*A3) function in the "diagnostics"
(*A) function group.
"
Caution!
A reset impairs the measurement. You should perform a basic setup again following a reset.
All the calibration data are deleted after a reset. A complete calibration is needed to put the
measurement back into operation.
!
Note!
The default values of each parameter are shown in bold in the menu overview in the appendix.