Gammapilot M FMG60 with FOUNDATION Fieldbus Operation
Endress+Hauser 43
5.6.4 Block model of the Gammapilot M
The Gammapilot M contains the following blocks:
Resource block (RB2)
See Operating Instructions BA 013S: "Foundation Fieldbus - Overview"
Transducer block (TB)
Contains the parameters relevant to measurement
Diagnostic block (DIAG)
Contains the diagnostic parameters of the Gammapilot M
Display block (DISP)
contains the configuration parameters for the display module VU331
Analog Input block 1 or 2 (AI)
Scale the signals of the Transducer block and transmit them to the PLC
PID block (PID)
See Operating Instructions BA013S: "Foundation Fieldbus - Overview"
Arithmetic block (AR)
See Operating Instructions BA013S: "Foundation Fieldbus - Overview"
Input Selector block (IS)
See Operating Instructions BA013S: "Foundation Fieldbus - Overview"
Signal Characterizer block (SC)
See Operating Instructions BA013S: "Foundation Fieldbus - Overview"
Integrator block (IT)
See Operating Instructions BA013S: "Foundation Fieldbus - Overview"
Default block configuration
The input and output variables of the blocks can be interconnected by a network configuration tool
(e.g. NI-Fieldbus configurator). The figure below shows, how these connections are set by default.
L00-FMG60xxx-02-00-00-en-001
OUT
OUT
Physical Block
parameters of the
physical unit, e.g.
tag No.
Display
Sensor
signal evaluation
Transducer Block
Parameters that
describe the
device (calibration,
linearisationetc.)
Analog Input Function Block 1
parameters that are importantto
the process control system,
e.g. scaling, status
Analog Input Function Block 2
parameters that are importantto
the p
rocess control system,
e.g. scaling, status
PID Block
Automation
functions
Primary value
(mainvalue)
Secondary value
(pulse rate)
Third value
(temperature)