Commissioning Gammapilot M FMG60 with FOUNDATION Fieldbus
76 Endress+Hauser
"Avg. pulse rate" (*11)
The average pulse rate is displayed in this function (after selection of "start" in the previous function).
Initially, this value fluctuates (because of the decay statistics), but due to the integration it reaches
an average value in the course of time. The longer the averaging is performed the lower are the
remaining fluctuations.
L00-FMG60xxx-05-00-00-xx-023
Initially, the pulse rate strongly fluctuates. In the course of time an average value is reached.
If the value is sufficiently stable, the function can be left by pressing "E".
Thereafter, the Gammapilot M changes to the "calibration" (*15) function. Select "stop/edit"
to stop the averaging procedure. The value is then automatically transmitted to the "full calibr."
(*16) or "empty calibr." (*18) function respectively.
!
Note!
The maximum integration time is 1000 s. The value is then transmitted automatically to the
"full calibr." (*16) or "empty calibr." (*18) function.
!
Note!
The integration is not terminated by pressing "E" in the "avg. pulse rate" (*11) function. It is
continued until the selection of "stop/edit" in the "calibration" (*15) function. This may result
in a slight deviation between the last displayed average pulse rate and the final "full calibr." (*16)
or "empty calibr. (*18).
⇒
avg. pulse rate *11
2548 cps
ENDRESS + HAUSER
E
+
–
c/s
t
"E"