Gammapilot M FMG60 with FOUNDATION Fieldbus Commissioning
Endress+Hauser 77
"Full calibration" (*16)
"Empty calibration" (*18)
The pulse rate of the respective calibration point is displayed in these functions. The displayed value
must be confirmed by pressing "E".
"-1" indicates, that no background calibration is present yet. In this case, there are two options:
either return to the "calibration" (*15) function and restart the calibration
or enter a known or calculated pulse rate (manual calibration).
"Next point" (*19)
This function is used to specify, if a further calibration point is to be entered or not.
Options:
no
This option must be selected after both calibration points have been entered. After this selection
the Gammapilot M returns to the group selection and the calibration is completed.
yes
This option must be selected if only one calibration point has yet been entered. After this selection
the Gammapilot M returns to the "calibr. point" (*13) function and the next point can be
entered.
6.4.4 Additional settings
After the calibration, the Gammapilot M outputs the measured value via the Foundation Fieldbus
interface.
Many additional functions are available for optimizing the measuring point. They can be configured
as required. A detailed description of all instrument functions is given in Operating Instructions
BA287F, "Gammapilot M - Description of Device Functions". This document can be found on the
CD-ROM supplied for the "ToF Tool - FieldTool Package".
full calibr. *16
33 cps
ENDRESS + HAUSER
E
+
empty calibr. *18
2548 cps
ENDRESS + HAUSER
E
+
next point *19
Âno
yes
ENDRESS + HAUSER
E
+