Gammapilot M FMG60 with FOUNDATION Fieldbus Troubleshooting
Endress+Hauser 95
9.1.4 Error codes
Code Description Remedy
A102 checksum error call Endress+Hauser Service
W103 initializing wait for completion of the initializing procedure
A106 downloading wait for completion of the download
A110 checksum error call Endress+Hauser Service
A111 electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
A113 electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
A114 electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
A116 download error repeat download
A121 electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
W153 initializing wait for completion of the initializing procedure
A160 checksum error call Endress+Hauser Service
A165 electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
A291 slave error check basic setup and connection of the slave transmitter
A503 wrong sensor type call Endress+Hauser Service
W513 calibration integration running Wait, until a stable pulse rate has been reached; then, terminate
integration (by pressing "E" in the "average pulse rate" (*11)
function)
W514 PT-100 calibration wait for the calibration to complete.
if error still persists: call Endress+Hauser Service
A531 sensor electronics defect switch device off/on;
if error still is present: call Endress+Hauser service or exchange
transmitter
A532 sensor voltage error call Endress+Hauser Service
A533 wrong sensor software version call Endress+Hauser Service
A535 sensor regulation error call Endress+Hauser Service
W536 high voltage near limit call Endress+Hauser Service
A538 sensor communication error call Endress+Hauser Service
A602 linearization table not plausible check monotony of the linearization table; if necessary, adjust table
("linearization" (*4) function group)
A612 linearization table not defined enter or complete linearization table
( "linearization" (*4) function group)
The linearization table must contain the end points 0% = 1000 cps
(standardized) and 100% = 0 cps (standardized).
When entering via ToF-Tool:
Select the right type of table (lin. tab. "level" or "concentration")
W621 simulation on Switch off linearization
("output" (*6) function group
"simulation" (*65) function)
W640 SIL lock device SIL_locking not completed
W642 I_back calibration running calibration of the current readback path active