Diagnostics and troubleshooting Proline t-mass I 300 HART
150 Endress+Hauser
11.10 Event logbook
11.10.1 Reading out the event logbook
A chronological overview of the event messages that have occurred is provided in the
Events list submenu.
Navigation path
Diagnostics menu → Event logbook submenu → Event list
F
I1091Config.change
I1157Mem.err.ev.list
F311Electr.failure
/../Eventlist
0d01h19m10s
A0014008-EN
 43 Taking the example of the local display
A maximum of 20 event messages can be displayed in chronological order.
If the Extended HistoROM application package (order option) is enabled in the device,
the event list can contain up to 100 entries .
The event history includes entries for:
Diagnostic events →  145
Information events →  151
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
Diagnostic event
: Occurrence of the event
: End of the event
Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
Via local display →  141
Via web browser →  142
Via "FieldCare" operating tool →  144
Via "DeviceCare" operating tool →  144
For filtering the displayed event messages →  150
11.10.2 Filtering the event logbook
Using the Filter options parameter you can define which category of event message is
displayed in the Events list submenu.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
All
Failure (F)
Function check (C)
Out of specification (S)
Maintenance required (M)
Information (I)