Diagnostics and troubleshooting Proline t-mass I 300 HART
150 Endress+Hauser
11.10 Event logbook
11.10.1 Reading out the event logbook
A chronological overview of the event messages that have occurred is provided in the
Events list submenu.
Navigation path
Diagnostics menu → Event logbook submenu → Event list
F
I1091Config.change
I1157Mem.err.ev.list
F311Electr.failure
/../Eventlist
0d01h19m10s
A0014008-EN
43 Taking the example of the local display
• A maximum of 20 event messages can be displayed in chronological order.
• If the Extended HistoROM application package (order option) is enabled in the device,
the event list can contain up to 100 entries .
The event history includes entries for:
• Diagnostic events → 145
• Information events → 151
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
• : Occurrence of the event
• : End of the event
• Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
• Via local display → 141
• Via web browser → 142
• Via "FieldCare" operating tool → 144
• Via "DeviceCare" operating tool → 144
For filtering the displayed event messages → 150
11.10.2 Filtering the event logbook
Using the Filter options parameter you can define which category of event message is
displayed in the Events list submenu.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)