iTEMP TMT142B Operating menu and parameter description
Endress+Hauser 67
Description Displays the diagnostic messages that occurred in the past. The last 10 messages are listed
in chronological order.
User interface Symbol for event behavior and diagnostic event.
Additional information Example for display format:
F201-Electronics faulty
Previous diag n channel
Navigation Diagnostics → Event logbook → Previous diag n channel
Description Displays the function module to which the diagnostic message refers.
User interface Device
Sensor
Device temperature
Current output
Sensor RJ
Time stamp n
Navigation Diagnostics → Event logbook → Time stamp n
Description Displays the time stamp of the current diagnostic message in relation to the operating
time.
User interface Hours (h)
14.1.4 Submenu: Simulation
Diagnostic event simulation
Navigation Diagnostics → Simulation → Diagnostic event simulation
Description Switches diagnostic simulation on and off. The status signal indicates a category "C"
diagnostic message ("function check") while the simulation is running.
Options Enter one of the diagnostic events using the dropdown menu →  40. The assigned
status signals and diagnostic behaviors are used in the simulation mode. Select 'Off' to quit
the simulation.
Example: x043 Short circuit
Factory setting Off