TMT162
16 Endress+Hauser
Certificates and approvals
CE-Mark The device meets the legal requirements of the EC directives. Endress+Hauser confirms that the device has
been successfully tested by applying the CE mark.
ATEX
FM (Factory Mutual) Intrinsic Safety; Non-Incendive
Mark:
• IS / I / 1 / ABCD / T4 Ta = 85 °C, Entity;
• NI / I / 2 / ABCD / T4 Ta = 85 °C;
• NI / I / 2 / ABCD / T4 Ta = 85 °C, NIFW;
• I / 0 / AEx ia IIC T4 Ta = 85 °C, Entity;
For Entity Parameters/Non-incendive Field Wiring parameters, refer to CSA approval under: Intrinsical Safety.
Explosion-proof
1)
, Dust ignition-proof
Mark:
• XP / I / 1 / ABCD / T6 Ta = 55 °C; T5 Ta = 70 °C; T4 Ta = 85 °C
• DIP / II, III / 1 / EFG / T6 Ta = 55 °C; T5 Ta = 70 °C; T4 Ta = 85 °C
• XP / I / 1 / IIC T6
For connection data, see CSA approval under: Explosion-proof, Dust ignition-proof.
Ta = 55 °C; T5 Ta = 70 °C; T4 Ta = 85 °C; Type 4X; IP66, IP67
ATEX II1G Ex ia IIC T6/T5/T4 HART
®
FOUNDATION Fieldbus™
PROFIBUS
®
PA
Power supply (+ and - terminals) U
i
≤ 30 V DC
I
i
≤ 300 mA
P
i
≤ 1000 mW
C
i
≤ 5 nF
L
i
≈ 0
U
i
≤ 17.5 V DC or:
I
i
≤ 500 mA
P
i
≤ 5.5 W
C
i
≤ 5 nF
L
i
= 10 μH
U
i
≤ 24 V DC
I
i
≤ 250 mA
P
i
≤ 1.2 W
Suitable for connecting to a fieldbus system in accordance
with the FISCO/FNICO model (valid for FOUNDATION
Fieldbus™ protocol)
ATEX II3G Ex nA II T6/T5/T4
HART
®
FOUNDATION
Fieldbus™
PROFIBUS
®
PA
Power supply (+ and - terminals) U ≤ 40 V DC U ≤ 32 V DC
Output I = 4 to 20 mA Current cons. I ≤ 12 mA Current cons. I ≤ 11 mA
ATEX II3G Ex nL IIC T6/T5/T4
HART
®
FOUNDATION Fieldbus™
PROFIBUS
®
PA
Power supply (+ and - terminals) - U
i
≤ 32 V DC
C
i
≤ 5 nF
L
i
= 10 μH
Temperature range
with display
without display
T6
T5
T4
T4
Ta = -40 °C to +55 °C
Ta = -40 °C to +70 °C
Ta = -40 °C to +70 °C
Ta = -40 °C to +85 °C
ATEX II2D Ex tD A21 IP67 T110°C
ATEX II2G Ex d IIC T6/T5/T4
HART
®
FOUNDATION Fieldbus™
PROFIBUS
®
PA
Power supply (+ and - terminals) U ≤ 40 V DC
P ≤ 3 W
U ≤ 35 V DC
P ≤ 3 W
Temperature range for Ex d T6
T5
T4
Ta = -40 °C to +55 °C
Ta = -40 °C to +70 °C
Ta = -40 °C to +80 °C
Temperature range for dust Ta = -40 °C to +80 °C
1) not available for T17 housing