TMT162
Endress+Hauser 19
Ordering information
Product structure
TMT162 iTEMP
®
TMT162, Field transmitter, Temperature transmitter; Application: RTD, TC, Ohm, mV; 2-wire 4...20 mA; galvanic
isolation; NEMA 4x, IP67; Dual compartment; GL (German Lloyd) Marine (HART
®
device only); UL listed
Approval:
A Non-hazardous area
B ATEX II1G EEx ia IIC T4/T5/T6
C FM IS, NI I/1+2/A-D
D CSA IS, NI I/1+2/A-D
E ATEX II2G EEx d IIC T6
F FM XP, DIP I,II,III/1+2/A-G
G CSA XP, DIP I,II,III/1+2/A-G
H ATEX EEx d, EEx ia
J FM XP, DIP, IS, NI I,II,III/1+2/A-G
K CSA XP, DIP, IS, NI I,II,III/1+2/A-G
L ATEX II3G EEx nA IIC T4/T5/T6
M FM + CSA XP, DIP, IS, NI, I,II,III/1+2/A-G
N ATEX II1/2D
O CSA General Purpose
P IECEx Ex ia IIC T6/T5/T4
R IECEx Ex d IIC T6/T5/T4
S IECEx Ex ia Ex d IIC T6/T5/T4
T ATEX II1/2GD EEx ia IIC T4/T5/T6
1 NEPSI Ex ia IIC T4-T6
2 NEPSI Ex nA II T4-T6
3 NEPSI Ex d IIC T4-T6
4 TIIS Ex ia IIC T4
5 TIIS Ex ia IIC T6
6 TIIS Ex d IIC T6
Housing:
1 Alu, w/o display
2 Alu + display, illum.
3 316L, w/o display
4 316L + display, illum.
5 T17, 316L, w/o display
6 T17, 316L + display, illum.
Cable Entry:
A 2x thread NPT1/2, 1x plug M12 PA
1)
1) Fieldbus connector is pre-assembled and wired at factory
B 2x thread M20x1.5, 1x plug M12 PA
1)
, 1x gland M20
C 2x thread NPT1/2 1x plug 7/8" FF
1)
D 2x thread M20x1.5, 1x plug 7/8" FF
1)
, 1x gland M20
E 2x thread M20x1.5, 1x plug 7/8" PA
1)
, 1x gland M20
1 2x thread NPT ½“
2 2x thread M20x1.5
4 2x thread G½“
5 1x thread M24x1.5 + 1x M20x1.5
6 2x gland M20x1.5
Mounting Bracket:
1 Not selected
2 Wall / pipe 2", L-shape, 304
3 Pipe 2", U-shape, 316L
Configuration Connection:
A Factory setup
1 Thermocouple TC
2 RTD 2-wire
3 RTD 3-wire
4 RTD 4-wire
Configuration Sensor Type:
A Factory setup
1 Pt100 (-200 to 850 °C, -328 to 1562 °F, min. span 10 K, 18 °F), as per IEC 60751
2 Ni100 (-60 to 250 °C, -76 to 482 °F, min. span 10 K, 18 °F)
3 Pt500 (-200 to 250 °C, -328 to 482 °F, min. span 10 K, 18 °F)
4 Pt100 (-200 to 649 °C, -328 to 1200 °F, min. span 10 K, 18 °F) as per JIS C1604-81