EasyManua.ls Logo

Endress+Hauser iTEMP TMT162 - Page 4

Endress+Hauser iTEMP TMT162
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TMT162
4 Endress+Hauser
System integration via PROFIBUS
®
PA
a0010723-en
Input
Measured variable Temperature (temperature linear transmission behavior), resistance and voltage.
Measuring range The transmitter records different measuring ranges depending on the sensor connection and input signals.
Type of input
0-10bar
0-10bar
PROFIBUS-DP
 
PROFIBUS-PA
IEC 61158-2
31.25 kbit/s
°C
TMT162
°C
TMT162
ENDRESS+ HAUSER
ENDRESS+ HAUSER
Visualization and monitoring
e. g. P View, FieldCare
and diagnostic software
Up to 32 devices per segment
PLC
segment
coupler
FISCO
segment
coupler
segment
coupler
Type of input Designation Measuring range limits Min. span
Resistance thermometer (RTD)
to IEC 60751
(α = 0.00385)
to JIS C1604-81
(α = 0.003916)
to DIN 43760
(α = 0.006180)
to Edison Copper Winding No.15
(α = 0.004274)
to Edison Curve
(α = 0.006720)
to GOST
(α = 0.003911)
to GOST
(α = 0.004278)
Pt100
Pt200
Pt500
Pt1000
Pt100
Ni100
Ni1000
Cu10
Ni120
Pt50
Pt100
Cu50, Cu100
Pt100 (Callendar - van Dusen)
Nickel polynomial (only PROFIBUS
®
PA)
Copper polynomial (only PROFIBUS
®
PA)
-200 to 850 °C (-328 to 1562 °F)
-200 to 850 °C (-328 to 1562 °F)
-200 to 250 °C (-328 to 482 °F)
-200 to 250 °C (-328 to 482 °F)
-200 to 649 °C (-328 to 1200 °F)
-60 to 250 °C (-76 to 482 °F)
-60 to 150 °C (-76 to 302 °F)
-100 to 260 °C (-148 to 500 °F)
-70 to 270 °C (-94 to 518 °F)
-200 to 1100 °C (-328 to 2012 °F)
-200 to 850 °C (-328 to 1562 °F)
-200 to 200 °C (-328 to 392 °F)
10 to 400 Ω
10 to 2000 Ω
10 to 400 Ω
10 to 2000 Ω
10 to 400 Ω
10 to 2000 Ω
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 °C (18 °F)
10 Ω
100 Ω
10 Ω
100 Ω
10 Ω
100 Ω
Type of connection: 2-wire, 3-wire or 4-wire connection, sensor current: 0.3 mA
With 2-wire circuit, compensation of wire resistance possible (0 to 30 Ω)
With 3-wire and 4-wire connection, sensor wire resistance to max. 50 Ω per wire