TMT162 FF/PA Safety instructions
Endress+Hauser 7
Safety instructions
iTEMP
®
TMT162 PROFIBUS
®
PA-, FOUNDATION Fieldbus™ protocol
Temperature field transmitter
for electrical apparatus certified for use in explosion-hazardous areas
Areas of application:
0
Designation according to Directive 94/9/EG:
4 0
II 1/2 or 2G
Equipment Group II
Equipment category: Sensor
category 1/Housing category 2
or equipment category 2
For explosive mixtures of air and
combustible gases, vapours or mists
Equipment Category Explosive gas-air mixtures (G)
Category 1 Zone 0, 1 or 2
Category 2 Zone 1 or 2
Category 3 Zone 2
Designation of explosion protection: EEx d IIC T6/T5/T4
Electrical apparatus with explosion protection to
European standard
Type of protection
Apparatus group
Temperature class
XA058R/09/a3
71003981
II 1/2G or 2G
Associated Documentation
Operating instructions: BA224R/09/en (FF-protocol)
Operating instructions: BA275R/09/en (PA-protocol)
Technical information: TI086R/09/en