iTEMP TMT162 Operating menu and parameter description
Endress+Hauser 103
Factory setting 0x11CE
Device revision
Navigation Expert → Communication → HART info → Device revision
Description Use this function to view the device revision with which the device is registered with the
HART
®
FieldComm Group. It is needed to assign the appropriate device description file
(DD) to the device.
Display 4
Factory setting 4 (0x04)
Device ID
Navigation Expert → Communication → HART info → Device ID
Description A unique HART identifier is saved in the device ID and used by the control systems to
identify the device. The device ID is also transmitted in command 0. The device ID is
determined unambiguously from the serial number of the device.
Display ID generated for specific serial number
Manufacturer ID
Navigation Expert → Communication → HART info → Manufacturer ID
Expert → Diagnostics → Device information → Manufacturer ID
Description Use this function to view the manufacturer ID under which the device is registered with
the HART FieldComm Group.
Display 2-digit hexadecimal number
Factory setting 0x0011
HART revision
Navigation Expert → Communication → HART info → HART revision
Description Use this function to display the HART revision of the device.