Installation iTEMP TMT162
14 Endress+Hauser
4.3 Display mounting
1
2
3
4
90°
90°
90°
90°
3 mm
WRITE
LOCK
ON
OFF
Proof-
Test
A0025417
 4 4 display installation positions, attachable in 90° stages
1
Cover clamp
2 Housing cover with O-ring
3 Display with fitting kit and twist protection
4 Electronics module
1. Remove the cover clamp (1).
2. Unscrew the housing cover together with the O-ring (2).
3. Remove the display with twist protection (3) from the electronics module (4). Fit the
display with the fitting kit in the desired position in 90° stages and plug it into the
correct slot on the electronics module.
4. Then screw the housing cover together with the O-ring.
5. Fit the cover clamp (1) back on.
4.4 Post-installation check
After installing the device, always run the following final checks:
Device condition and specifications Notes
Is the device undamaged (visual inspection)? -
Do the ambient conditions match the device specification (e.g. ambient temperature,
measuring range, etc.)?
→  46