Diagnostics and troubleshooting iTEMP TMT162
38 Endress+Hauser
Diagnostic
number
Short text Corrective measure
Status
signal
from the
factory
Customizable
1)
Diagnosti
c
behavior
from the
factory
Customizable
2)
Not
customizable
Not
customizable
Diagnostics for the sensor
001 Device failure - sensor n
3)
(sensor RJ)
1. Restart device
2. Replace electronics
F Alarm
041 Sensor interrupted - sensor n 1. Check electrical wiring.
2. Replace sensor.
3. Check connection type.
F Alarm
042 Sensor n corroded 1. Check sensor.
2. Replace sensor.
M Warning
043 Short-circuit sensor n 1. Check electrical connection.
2. Check sensor.
3. Replace sensor or cable.
F Alarm
044 Sensor drift detected 1. Check sensor or main electronics.
2. Replace sensor or main electronics.
M Warning
047 Sensor limit reached sensor n
(sensor RJ)
1. Check sensor.
2. Check process conditions.
S Warning
048 Drift detection not possible 1. Check electrical connection.
2. Check sensor.
3. Replace sensor.
M Warning
062 Sensor connection faulty sensor
n (sensor RJ)
Check sensor connection. F Alarm
105 Calibration interval 1. Execute calibration and reset calibration
interval.
2. Switch off calibration counter.
M Warning
145 Compensation reference point
sensor n
1. Check terminal temperature.
2. Check external reference point.
F Alarm
Diagnostics for the electronics
201 Electronics faulty 1. Restart device.
2. Replace electronics.
F Alarm
221 Reference sensor defective
sensor RJ
Replace device. M Alarm
241 Firmware faulty 1. Restart device.
2. Power cycle device.
3. Replace electronics.
F Alarm
242 Firmware incompatible 1. Check firmware version.
2. Flash or replace main electronics.
F Alarm
261 Electronics module defective 1. Restart device.
2. Replace main electronics module.
F Alarm
283 Memory content inconsistent 1. Restart device.
2. Replace electronics.
F Alarm
286 Data storage inconsistent 1. Repeat safe parameterization.
2. Replace electronics.
F Alarm
Diagnostics for the configuration
401 Factory reset active Factory reset active, please wait. C Warning
402 Initialization active sensor n
(sensor RJ)
Initialization active, please wait. C Warning

Table of Contents