iTEMP TMT162 Technische Daten
Endress+Hauser 51
Maximum measured error In accordance with DIN EN 60770 and the reference conditions specified above. The
measured error data correspond to ±2 s (Gaussian distribution), i.e. 95.45%. The data
include non-linearities and repeatability.
Typical
Standard Designation Measuring range Typical measured error (±)
Resistance thermometer (RTD) as per standard Digital value
1)
Value at current
output
IEC 60751:2008 Pt100 (1)
0 to +200 °C (32 to +392 °F)
0.08 °C (0.14 °F) 0.1 °C (0.18 °F)
IEC 60751:2008 Pt1000 (4) 0.06 °C (0.11 °F) 0.1 °C (0.18 °F)
GOST 6651-94 Pt100 (9) 0.07 °C (0.13 °F) 0.09 °C (0.16 °F)
Thermocouples (TC) as per standard Digital value
Value at current
output
IEC 60584, Part 1 Type K (NiCr-Ni) (36)
0 to +800 °C (32 to +1 472 °F)
0.22 °C
(0.4 °F) 0.24 °C (0.43 °F)
IEC 60584, Part 1 Type S (PtRh10-Pt) (39) 1.17 °C (2.1 °F) 1.33 °C (2.4 °F)
GOST R8.8585-2001 Type L (NiCr-CuNi) (43) 2.0 °C (3.6 °F) 2.4 °C (4.32 °F)
1)
Measured value transmitted via HART
®
.
Measured error for resistance thermometers (RTD) and resistance transmitters
Standard Designation Measuring range Measured error (±)
Digital
1)
D/A
2)
Maximum
3)
Based on measured value
4)
IEC 60751:2008
Pt100 (1)
–200 to +850 °C
(–328 to +1 562 °F)
≤ 0.11 °C (0.2 °F)
ME = ± (0.06 °C (0.11 °F) +
0.005
% * (MV - LRV))
0.03 % (
4.8 µA)
Pt200 (2) ≤ 0.18 °C (0.32 °F)
ME = ± (0.05 °C (0.09 °F) +
0.012% * (MV - LRV))
Pt500 (3) –200 to +500 °C (–328 to +932 °F) ≤ 0.11 °C (0.2 °F)
ME = ± (0.03 °C (0.05 °F) +
0.012% * (MV - LRV))
Pt1000 (4) –200 to +250 °C (–328 to +482 °F) ≤ 0.07 °C (0.13 °F)
ME = ± (0.02 °C (0.04 °F) +
0.012% * (MV - LRV))
JIS C1604:1984 Pt100 (5) –200 to +510 °C (–328 to +950 °F) ≤ 0.09 °C (0.16 °F)
ME = ± (0.05 °C (0.09 °F) +
0.006% * (MV - LRV))
GOST 6651-94
Pt50 (8)
–185 to +1 100 °C
(–301 to +2 012 °F)
≤ 0.20 °C (0.36 °F)
ME = ± (0.1 °C (0.18 °F) +
0.008% * (MV - LRV))
Pt100 (9)
–200 to +850 °C
(–328 to +1 562 °F)
≤ 0.11 °C (0.2 °F)
ME = ± (0.05 °C (0.09 °F) +
0.006% * (MV - LRV))
DIN 43760 IPTS-68
Ni100 (6)
–60 to +250 °C (–76 to +482 °F) ≤ 0.05 °C (0.09 °F)
ME = ± (0.05 °C (0.09 °F) -
0.006% * (MV - LRV))
Ni120 (7)
OIML R84: 2003 /
GOST 6651-2009
Cu50 (10) –180 to +200 °C (–292 to +392 °F) ≤ 0.11 °C (0.2 °F)
ME = ± (0.10 °C (0.18 °F) +
0.006% * (MV - LRV))
Cu100 (11) –180 to +200 °C (–292 to +392 °F)
≤ 0.06 °C (0.11 °F)
ME = ± (0.05 °C (0.09 °F) +
0.003% * (MV - LRV))
Ni100 (12)
–60 to +180 °C (–76 to +356 °F)
ME = ± (0.06 °C (0.11 °F) -
0.005% * (MV - LRV))
Ni120 (13) ≤ 0.05 °C (0.09 °F)
ME = ± (0.05 °C (0.09 °F) -
0.005% * (MV - LRV))
OIML R84: 2003, GOST
6651-94
Cu50 (14) –50 to
+200 °C (–58 to +392 °F) ≤ 0.11 °C (0.2 °F)
ME = ± (0.1 °C (0.18 °F) +
0.004% * (MV - LRV))