Operating menu and parameter description iTEMP TMT162
76 Endress+Hauser
Description Alarm delay for drift detection monitoring.
Useful for example in the event of different thermal mass ratings for the sensors in
conjunction with a high temperature gradient in the process.
User entry 5 to 255 s
Factory setting 5 s
Drift/difference set point
Navigation Setup → Advanced setup → Sensor → Drift/difference set point
Expert → Sensor → Diagnostic settings → Drift/difference set point
Prerequisite The Drift/difference mode parameter must be activated with the Out band (drift) or In
band option.
Description Use this function to configure the maximum permissible measured value deviation
between sensor 1 and sensor 2 which results in drift/difference detection.
User entry 0.1 to 999.0 K (0.18 to 1
798.2 °F)
Factory setting 999.0
Sensor switch set point
Navigation Setup → Advanced setup → Sensor → Sensor switch set point
Expert → Sensor → Diagnostic settings → Sensor switch set point
Description Use this function to set the threshold value for sensor switching.→ 72
Additional information The threshold value is relevant if the sensor switching function is assigned to a HART
®
variable (PV, SV, TV, QV).
User entry Depends on the sensor types selected.
Factory setting 850 °C
"Current output" submenu
Adjustment of the analog output (4 and 20 mA current trimming)
Current trimming is used to compensate the analog output (D/A conversion). Here, the
output current of the transmitter must be adapted so that it suits the value expected at the
higher-level system.