iTEMP TMT162 Operating menu and parameter description
Endress+Hauser 79
Navigation Setup → Advanced setup → Current output → Reset trim
Expert → Output → Reset trim
Description The Wizard resets the 4 to 20 mA values for trimming to the default value.
User entry Activate the button
"Display" submenu
The settings for displaying the measured value on the optional display are made in the
"Display" menu.
These settings do not affect the output values of the transmitter, and are only used to
specify the display format on the screen.
Display interval
Navigation Setup → Advanced setup → Display → Display interval
Expert → System → Display → Display interval
Description Use this function to set the length of time the measured values are displayed if the values
alternate on the display. This type of change is only generated automatically if several
measured values are specified.
The Value 1 display - Value 3 display parameters are used to specify which
measured values are shown on the local display → 79.
User entry 4 to
20 s
Factory setting 4 s
Value 1 display (Value 2 or 3 display)
Navigation Setup → Advanced setup → Display → Value 1 display (Value 2 or 3 display)
Expert → System → Display → Value 1 display (Value 2 or 3 display)
Description Use this function to select one of the measured values to be shown on the local display.
User entry • Process value
• Sensor 1
• Sensor 2
• Output current
•
Percent of range
• Device temperature
Factory setting Process value