Operating menu and parameter description iTEMP TMT162
86 Endress+Hauser
14.2.2 "Event logbook" submenu
Previous diagnostics n
n = Number of diagnostics messages (n = 1 to 5)
Navigation Diagnostics → Diagnostic list → Previous diagnostics n
Expert → Diagnostics → Diagnostic list → Previous diagnostics n
Description Use this function to display the diagnostics messages that occurred in the past. The last 5
messages are listed in chronological order.
Display Symbol for event behavior and diagnostic event.
Additional information Example for display format:
F261-Electronics modules
Previous diag n channel
Navigation Diagnostics → Diagnostic list → Previous diag channel
Expert → Diagnostics → Diagnostic list → Previous diag channel
Description Use this function to display the possible sensor input to which the diagnostics message
refers.
Display - - - - - -
Sensor 1
Sensor 2
Device temperature
Current output
Terminal temperature
14.2.3 "Device information" submenu
Device tag
Navigation Setup → Device tag
Diagnostics → Device information → Device tag
Expert → Diagnostics → Device information → Device tag
Description Use this function to enter a unique name for the measuring point so it can be identified
quickly within the plant. This name is shown on the display. →
 23
User entry Max. 32 characters such as letters, numbers or special characters (e.g. @, %, /)
Factory setting 32 x '?'