iTEMP TMT162 Operating menu and parameter description
Endress+Hauser 99
Drift/difference mode → 75
Navigation Setup → Advanced setup → Sensor → Drift/difference mode
Expert → Sensor → Diagnostic settings → Drift/difference mode
Drift/difference alarm delay → 75
Navigation Setup → Advanced setup → Sensor → Drift/difference alarm delay
Expert → Sensor → Diagnostic settings → Drift/difference alarm delay
Drift/difference set point → 76
Navigation Setup → Advanced setup → Sensor → Drift/difference set point
Expert → Sensor → Diagnostic settings → Drift/difference set point
Control
Navigation Expert → Sensor → Diagnostic settings → Control
Description Option to control the calibration counter.
The countdown duration (in days) is specified with the Start value parameter.
Options • Off: Stops the calibration counter
• On: Starts the calibration counter
• Reset + run: Resets to the set start value and starts the calibration counter
Factory setting Off
Start value
Navigation Expert → Sensor → Diagnostic settings → Start value
Description Use this function to set the start value for the calibration counter.
User entry 0 to 1826 d (days)
Factory setting 1826