iTEMP TMT162 Operating menu and parameter description
Endress+Hauser 99
Drift/difference mode →  75
Navigation Setup → Advanced setup → Sensor → Drift/difference mode
Expert → Sensor → Diagnostic settings → Drift/difference mode
Drift/difference alarm delay →  75
Navigation Setup → Advanced setup → Sensor → Drift/difference alarm delay
Expert → Sensor → Diagnostic settings → Drift/difference alarm delay
Drift/difference set point →  76
Navigation Setup → Advanced setup → Sensor → Drift/difference set point
Expert → Sensor → Diagnostic settings → Drift/difference set point
Control
Navigation Expert → Sensor → Diagnostic settings → Control
Description Option to control the calibration counter.
The countdown duration (in days) is specified with the Start value parameter.
Options Off: Stops the calibration counter
On: Starts the calibration counter
Reset + run: Resets to the set start value and starts the calibration counter
Factory setting Off
Start value
Navigation Expert → Sensor → Diagnostic settings → Start value
Description Use this function to set the start value for the calibration counter.
User entry 0 to 1826 d (days)
Factory setting 1826