EasyManua.ls Logo

Endress+Hauser iTEMP TMT71 - Page 52

Endress+Hauser iTEMP TMT71
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DRAFT
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
DRAFT DRAFT DRAFT
Technical data iTEMP TMT71/72
Operating influences The measured error data correspond to 2 σ (Gaussian distribution).
Influence of ambient temperature and supply voltage on operation for resistance thermometers (RTD) and resistance
transmitters
Designation Standard
Ambient temperature:
Influence (±) per 1 °C (1.8 °F) change
Supply voltage:
Influence (±) per V change
Digital
1)
D/A
2)
Digital D/A
Maximum Based on measured value Maximum Based on measured value
Pt100 (1)
Pt200 (2)
Pt500 (3)
Pt1000 (4)
IEC
60751:2008
0.013 °C
(0.023 °F)
0.017 °C
(0.031 °F)
0.008 °C
(0.014 °F)
0.005 °C
(0.009 °F)
0.0013% * (MV - LRV),
at least 0.003 °C (0.005 °F)
-
0.0013% * (MV - LRV),
at least 0.006 °C (0.011 °F)
- 0.003 %
0.007 °C
(0.013 °F)
0.009 °C
(0.016 °F)
0.004 °C
(0.007 °F)
0.003 °C
(0.005 °F)
0.007% * (MV - LRV),
at least 0.003 °C (0.005 °F)
-
0.007% * (MV - LRV),
at least 0.006 °C (0.011 °F)
- 0.003 %
Pt100 (5) JIS C1604:1984
Pt50 (8)
GOST 6651-94
0.009 °C
(0.016 °F)
0.017 °C
(0.031 °F)
0.0013% * (MV - LRV),
at least 0.003 °C (0.005 °F)
0.0015% * (MV - LRV),
at least 0.01 °C (0.018 °F)
0.004 °C
(0.007 °F)
0.009 °C
(0.016 °F)
0.007% * (MV - LRV),
at least 0.003 °C (0.005 °F)
0.007% * (MV - LRV),
at least 0.01 °C (0.018 °F)
Pt100 (9)
Ni100 (6)
Ni120 (7)
Cu50 (10)
Cu100 (11)
Ni100 (12)
Ni120 (13)
Cu50 (14)
DIN 43760
IPTS-68
OIML R84:
2003 /
GOST
6651-2009
OIML R84:
2003 /
GOST 6651-94
0.013 °C
(0.023 °F)
0.003 °C
(0.005 °F)
0.005 °C
(0.009 °F)
0.004 °C
(0.007 °F)
0.003 °C
(0.005 °F)
0.005 °C
(0.009 °F)
0.0013% * (MV - LRV),
at least 0.003 °C (0.005 °F)
-
-
-
-
-
-
-
0.007 °C
(0.013 °F)
0.001 °C
(0.002 °F)
0.005 °C
(0.009 °F)
0.004 °C
0.003 % (0.007 °F)
0.003 °C
(0.005 °F)
0.005 °C
(0.009 °F)
0.007% * (MV - LRV),
at least 0.003 °C (0.005 °F)
-
-
-
-
-
-
-
0.003 %
Resistance transmitter ()
10 to 400 4 m
0.001% * MV,
at least 1 m
0.003 %
2 m
0.0005% * MV,
at least 1 m
0.003 %
at least 15 m at least 5 m
1)
Measured value transmitted via HART
®
.
2) Percentages based on the configured span of the analog output signal
Influence of ambient temperature and supply voltage on operation for thermocouples (TC) and voltage transmitters
Designation Standard
Ambient temperature:
Influence (±) per 1 °C (1.8 °F) change
Supply voltage:
Influence (±) per V change
Digital
1)
D/A
2)
Digital D/A
Maximum Based on measured value Maximum Based on measured value
Type A (30)
Type B (31)
IEC 60584-1
0.07 °C
(0.126 °F)
0.04 °C
(0.072 °F)
0.003% * (MV - LRV),
at least 0.013 °C (0.023 °F)
-
0.003 %
0.03 °C
(0.054 °F)
0.02 °C
(0.036 °F)
0.0012% * (MV - LRV),
at least 0.013 °C (0.023 °F)
-
0.003 %
52 V. 1, Rev. 2, 4-10-2018 Endress+Hauser
10 to 2 000 20 m 0.001% * MV,
10 m 0.0005% * MV,

Table of Contents