EasyManua.ls Logo

Endress+Hauser iTEMP TMT82 - Connecting the Sensor Cables

Endress+Hauser iTEMP TMT82
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connection iTEMP TMT82
18 Endress+Hauser
3
3
3
3
3
3
1
1
1
1
1
1
1
4
4
2
2
2
2
2
RTD
RTD
RTD
Ω
5
5
6
6
6
RTD
RTD
6
4
4
4
4
TC
Ω Ω
Ω
Ω
+
-
1
2
3
4
5
6
-
+
3
S2
-
+
-
3
4
S1
1
red
(black)
white
red
red
4-wire
2-wire 3-wire
Bus connection and
supply voltage
Sensor 1
Sensor 2
Sensor 1
Sensor 2
white
red
red
red
white
white
red
(black)
white
(yellow)
white
(yellow)
A0042369-EN
 10 Terminal assignment of the field mount housing with separate terminal compartment
1 Fixed connection of the external reference junction, terminals 4, 5 and 6 (Pt100, IEC 60751, class B,
3-wire). It is not possible to connect a second thermocouple (TC) on sensor 2.