iTEMP TMT82 Operating options
Endress+Hauser 25
6 Operating options
6.1 Measured value display and operating elements
6.1.1 Option: Display TID10 with transmitter
The display may also be subsequently
ordered at any time after purchasing the
transmitter, see the 'Accessories' section in
the Operating Instructions for the device.
A0010227
16 Attach the display to the transmitter
6.1.2 Display elements
DIN rail transmitter
The DIN rail transmitter version does not have an interface to the LC display and
therefore does not have a local display either.
Two LEDs on the front indicate the device status.
Type Function and characteristic
Status LED (red) When the device is operating without errors, the device status is displayed.
This function can no longer be guaranteed in the event of an error.
• LED off: without diagnostic message
• LED is lit: diagnostics display, category F
• LED flashing: diagnostics display of categories C, S or M
Power LED (green) 'ON' When the device is operating without errors, the operating status is displayed.
This function can no longer be guaranteed in the event of an error.
• LED off: Power failure or insufficient supply voltage
• LED is lit: Supply voltage is OK (either via CDI or via supply voltage,
terminals 1+, 2-)
6.1.3 Local operation
NOTICE
‣
ESD - electrostatic discharge. Protect the terminals from electrostatic discharge. Failure
to observe this may result in the destruction or malfunction of parts of the electronics.