iTEMP TMT82 Diagnostics and troubleshooting
Endress+Hauser 39
11.2 Diagnostics events
11.2.1 Displaying diagnostics events
A0014837
A Display in the event of a warning
B Display in the event of an alarm
1 Status signal in the header
2 The display alternates between the primary measured value and the status - indicated by the appropriate
letter (M, C or S) - plus the defined error number.
3 The display alternates between "- - - -" (no valid measured value) and the status - indicated by the appropriate
letter (F) - plus the defined error number.
Status signals
Symbol Event
category
Meaning
Operating
error
An operating error has occurred. The measured value is no longer valid.
Service mode The device is in service mode (e.g. during a simulation).
Out of
specification
The device is being operated outside its technical specifications (e.g. during warm-
up or cleaning processes).
Maintenance
required
Maintenance is required. The measured value is still valid.
Diagnostic behavior
Alarm Measurement is interrupted. The signal outputs assume the defined alarm
condition. A diagnostic message is generated (status signal F).
Warning The device continues to measure. A diagnostic message is generated (status
signals M, C or S).