Levelflex FMP55 HART Electrical connection
Endress+Hauser 25
2-wire: 4-20mA HART, 4-20mA
1
3
+
+
2
4
4...20mA
HART
4...20mA
10 mm
Spare part
71108xxx
2- wire level
4-20 mA 4-20 mA
HART
[04/05]
open
-
-
11
A
1
3
+
+
2
4
4-20mA/
FIELDBUS
4-20mA/
2-channel overvoltage protection
-
-
[17]
14
13
12
+
+
+
+
-
-
-
-
1
1
2
2
3
3
9
9
5
5
8
8
6
6
7
7
4
4
4
4
+
+
+
+
4...20 mA
4...20 mA
10
B
4...20 mA
4...20 mA
"
"
"
"
#
A0013923
 8 Terminal assignment 2-wire, 4-20 mA HART, 4...20mA
A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection current output 2
2 Connection current output 1
3 Supply voltage for current output 1 (e.g. RN221N); Observe terminal voltage
4 Cable screen; observe cable specification
5 HART communication resistor (≥ 250 Ω): Observe maximum load
6 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)