EasyManua.ls Logo

Endress+Hauser Liquicap M FMI51 HART - Function: Basic Setup

Endress+Hauser Liquicap M FMI51 HART
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Liquicap M FMI51 HART
46 Endress+Hauser
Basic setup
function: Basic setup
subfunction:
- Medium property: no buildup, buildup
- Cal. type: Dry, Wet
function: Medium property
3)
subfunction: Medium property
- Conductive
- Nonconductive
4)
- interface
- unknown
subfunction: DC value
5)
- Value
subfunction: Unit level
6)
- % (percentage)
- m
- mm
- ft
- inch
function: Empty calibr.
subfunction: Value empty
- 0 %
subfunction: Measure capacity
- xxxx pF
subfunction: Confirm cal.:
- Yes
function: Full calibr.
subfunction: Value full
- 100%
subfunction: Measure capacity
- xxxx pF
subfunction: Confirm cal.:
- Yes
function: Output damping
subfunction: Output damping
- 1 s
7.3.2 Function: "Basic setup"
Subfunction: "Medium property"
The "No buildup" operating mode should be set for media that do not tend to form buildup
on the probe rod, like water, beverages, etc. For conductivity of 100 µS/cm, the measured
value is independent
7)
of the conductivity of the liquid.
In the "Buildup" operating mode, the buildup compensation function integrated with the
software is activated. In this operating mode, the measured value is independent
7)
of the
conductivity of the liquid conductivity of 1 000 µS/cm.
This compensates measuring errors caused by conductive media sticking to the probe rod
such as yogurt. This corresponds to buildup compensation.
3) The function is displayed if the function value "Dry" was selected under the subfunction "Cal. type".
4) The function value can be selected only for probes with a ground tube.
5) The subfunction is displayed if the function value "Nonconductive" was selected under the subfunction "Medium property".
6) The subfunction is displayed if the function value "Nonconductive" or "Conductive" was selected under the subfunction "Medium property".
7) Independent of concentration fluctuations.

Table of Contents