EasyManua.ls Logo

Endress+Hauser Liquicap M FMI51 - Zone 22 Application:

Endress+Hauser Liquicap M FMI51
28 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquicap M EG 05 022 / XA00346F-C
18 Endress+Hauser
Zone 22 Application: For the connetion to non-intrinsically safe circuits.
Electronic
insert
Type of protection/
level of protection
Electrical
data
Ambient temperature
electronic
Maximum surface
temperature (Ta max.)
Housing
ingress
protection
FEI50H II 3 D
Ex tD A22 IP6X
T 100 °C
Ui 30 V
Ii 120 mA
Pi 1 W
Ci 2.4 nF
Li 0
–50 °C Ta +60 °C +100 °C IP65
FEI57C II 3 D
Ex tD A22 IP6X
T 100 °C
Ui 19.2 V
Ii 108 mA
Pi 1 W
Ci 2.4 nF
Li 0
–50 °C Ta +60 °C +100 °C IP65
FEI55 II 3 D
Ex tD A22 IP6X
T 100 °C
Ui 36 V
Ii 100 mA
Pi 1 W
Ci 2.4 nF
Li 0
–50 °C Ta +60 °C +100 °C IP65
FEI57S II 3 D
Ex tD A22 IP6X
T 100 °C
Ui 16.1 V
Ii 100 mA
Pi 1 W
Ci 2.4 nF
Li 0
–50 °C
Ta +60 °C +100 °C IP65
FEI58 II 3 D
Ex tD A22 IP6X
T 100 °C
Ui 18 V
Ii 52 mA
Pi 170 mW
Ci 0
Li 0
–50 °C Ta +60 °C +100 °C IP65
Electronic
insert
Type of protection/
level of protection
Power supply Relay circuit
FEI52
(3 wire PNP)
II 3 D
Ex tD A22 IP6X
T 100 °C
10... 55 V DC
FEI54
(Relay)
II 3 D
Ex tD A22 IP6X
T 100 °C
19...253 V AC 253 V AC / 6 A *
1500 VA / cos ϕ = 1
750 VA / cos ϕ > 0.7
19... 55 V DC 30 V DC / 6 A *
125 V DC / 0.2 A
* for T13 housing: 4 A