Liquiline System CA80SI Technical data
Endress+Hauser 145
16 Technical data
16.1 Input
Measured values SIO
2
[mg/l, µg/l, ppm, ppb]
Measuring range
CA80SI-AAAH: 0.5 to 200 µg/l (ppb)
CA80SI-AAAJ: 50 to 5000 µg/l (ppb)
Types of input 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Analog current inputs (optional)
Input signal Depending on version
2 x 0/4 to 20 mA (optional), passive, potentially isolated
Current input, passive Span
> 0 to 20 mA
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V
Cable specification
(for optional sensors with
Memosens technology)
Cable type
Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12
round-pin connector (optional)
Cable length
Max. 100 m (330 ft)