Technical data Liquiline System CA80SI
150 Endress+Hauser
16.6 Power supply
Supply voltage The analyzer is fitted with a power cable and a safety plug with a cable length of 4.3
m (14.1 ft).
100 to 120 V AC / 200 to 240 V AC
50 or 60 Hz
NOTICE
The device does not have a power switch
The customer must provide a protected circuit breaker in the vicinity of the device.
The circuit breaker must be a switch or power switch, and you must label it as the
circuit breaker for the device.
Fieldbus connection Supply voltage: not applicable
Power consumption 1250 VA
Cable entries 4 x bores for M16, G3/8, NPT3/8", Memosens connection
4 x bores for M20, G1/2, NPT1/2"
Hose entries 4 x bores for M32 for sample inflow and outflow
Cable specification
Cable gland Permitted cable diameter
M16x1.5 mm 4 to 8 mm (0.16 to 0.32")
M12x1.5 mm (for order version M12 socket for Memosens sensors) 2 to 5 mm (0.08 to 0.20")
M20x1.5 mm 6 to 12 mm (0.24 to 0.48")
NPT³/₈" 4 to 8 mm (0.16 to 0.32")
G³/₈ 4 to 8 mm (0.16 to 0.32")
NPT½" 6 to 12 mm (0.24 to 0.48")
7 to 12 mm (0.28 to 0.48")
Cable glands mounted at the factory are tightened with 2 Nm.