Diagnostics and troubleshooting Liquiline CM442R/CM444R/CM448R
128 Endress+Hauser
11.12 Reset measuring instrument
11.13 Device information
11.13.1 System information
Diagnostics/Reset
Function Options Info
Device reset Options
•OK
•ESC
Restart and keep all the settings
Factory default Options
•OK
•ESC
Restart with factory settings
Settings that have not been saved are lost.
Diagnostics/System information
Function Options Info
Device tag Read only Individual device tag, "General settings"
Order code Read only You can order identical hardware with this code.
This code changes on account of changes to the
hardware and you can enter the new code you received
from the manufacturer here
1)
.
To find out what device version you have, enterthe order code in the search screen at the following address:
www.endress.com/order-ident
Orig. order code ext. Read only Complete order code for the original device, resulting
from the product structure.
Current order code ext. Free text Current code, taking into account changes to the
hardware. You must enter this yourself.
Serial number Read only The serial number allows you to access device data and
documentation on the Internet:
www.endress.com/device-viewer
Software version Read only Current version
HART
only with HART option
Read only
•Bus address
Unique address
Manufacturer ID
Device type
Device revision
Software revision
HART-specific information
The unique address is linked to the serial number and is
used to reach devices in a multi-drop environment.
Device and software version numbers are incremented
as soon as changes have been made.
Modbus
only with Modbus TCP or
Modbus RS485 option
Read only
•Enable
•Bus address
•Termination
•Modbus TCP Port 502
Modbus-specific information