Liquiline CM442/CM444/CM448 Technical data
Endress+Hauser 165
15.9 Power supply
15.9.1 Supply voltage
CM442
Depending on the version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V AC/DC +20 / -15 %, 50/60 Hz
CM444 and CM448
Depending on the version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V DC +20 / -15 %
15.9.2 Cable entries
Identification of the cable entry on the base of the housing Suitable gland
B, C, H, I, 1-8 M16x1.5 mm / NPT3/8" / G3/8
A, D, F, G M20x1.5 mm / NPT1/2" / G1/2
E-
) M12x1.5 mm
a0018025
Recommended assignment
1-8
A
B
C
D, F, G
H
I
E
Sensors 1-8
Power supply
RS485 In or M12 DP/RS485
Do not use
Current outputs and inputs, relays
Do not use
RS485 Out or M12 Ethernet
Do not use